=Description of the Experiment=
 
=Description of the Experiment=
[[File:Langmuir_with_plasma.jpg|318|thumb|Figure 1: Photo of the experimental setup]]
+
[[File:Langmuir_with_plasma.jpg|318|thumb|Fig. 1 - Photo of the experimental setup]]
  
Plasmas have different characteristics from other states of matter, and in order to measure them many diagnostic tools have been developed.  
+
Plasmas have different characteristics from other states of matter and many diagnostic tools have been developed in order to measure their parameters.  
This experiment allows the measurement of some of these characteristics using one of the most simple methods, the use of an electrical probe, the Langmuir Probe. It consists of a thin filament made of conductive material, placed inside the plasma, which either attracts or repels the electrons in the plasma according to its polarization. Measuring the probe I-V characteristic, that is, the relationship between the polarization voltage and the respective current going trough the it, one can take the electron temperature and density of the plasma.
+
This experiment allows the measurement of some of these characteristics using one of the most simple methods, an electrical probe, the Langmuir probe. It consists of a thin filament made of a conductive material placed inside the plasma, which either attracts or repels the electrons in the plasma in accordance with its polarization. Measuring the probe I-V characteristic, that is the relationship between the polarization voltage and the respective collected current, you can determine the electron temperature and density of the plasma.
  
 +
<table style="background-color:white"><tr><td style="vertical-align:top">
  
<div class="toccolours mw-collapsible mw-collapsed" style="width:420px">
+
<div class="toccolours mw-collapsible mw-collapsed" style="width:380px">
'''Ligações'''
+
'''Links'''
 
<div class="mw-collapsible-content">
 
<div class="mw-collapsible-content">
  
 
*Video: rtsp://elabmc.ist.utl.pt/langmuir.sdp
 
*Video: rtsp://elabmc.ist.utl.pt/langmuir.sdp
*Laboratory: New Experiments in elab1.ist.utl.pt[http://elab1.ist.utl.pt]
+
*Laboratory: Advanced in e-lab.ist.utl.pt[http://elab.ist.eu]
 
*Control room: Langmuir Probe
 
*Control room: Langmuir Probe
 
*Level: ****
 
*Level: ****
Line 17: Line 18:
 
</div>
 
</div>
 
</div>
 
</div>
 +
</td>
 +
<td style="vertical-align:top">
 +
 +
<div class="toccolours mw-collapsible mw-collapsed" style="width:380px">
 +
'''Videos'''
 +
<div class="mw-collapsible-content">
 +
{{#ev:youtube|EwK4Bn_7IVc|Glow discharche in Langmuir Probe experiment.|center}}
 +
 +
</div>
 +
</div>
 +
</td>
 +
</tr></table>
  
 
=Experimental Apparatus=
 
=Experimental Apparatus=
<!--[[File:condensador-seccao.png|thumb|Figure 2: Sectioned schematic of the capacitor, where a=12mm and b=16mm]]-->
+
<!--[[File:condensador-seccao.png|thumb|Fig. 2 Sectioned schematic of the capacitor, where a=12mm and b=16mm]]-->
  
The main chamber consists of a pirex cylinder on which a modified Filament Head is placed to function as a Langmuir probe.  
+
The vacuum chamber consists of a Pyrex cylinder on which a glow discharge is produced by RF power. A tungsten filament with a diameter of 0.2mm is placed in the middle of the chamber partially isolated by an alumina cover and only a 10mm tip is exposed to the plasma, acting as a Langmuir probe. This central filament has a ground reference around the probe with approximately 200mm, in order to be much longer than the probe itself.
  
The central filament of 0.2mm diameter has been covered with alumina tube leaving only a 1cm tip exposed to the plasma. The two outer filaments are connected to an "of the shelf" RF generator which ionizes the gas and generates the plasma. This is a simple 3.5W generator which outputs a 50kHz AC signal with 1kV amplitude. To improve stability and homogeneity of the electrical field two molybdenum plaques have been attached using low vapour pressure epoxy, to each outer filament.
+
The RF generator (3.5W, 50kHz, 1kV) that ionizes the gas and generates the plasma is connected to two parallel molybdenum plates similar to a capacitor.
 +
NOTE: Gas pressure correction should be computed according Pfeifer's vacuum gauge PPT100 spreadsheet.
  
 
=Protocol=
 
=Protocol=
Line 29: Line 43:
 
==Determination of the electron temperature and density==
 
==Determination of the electron temperature and density==
  
[[File:Langmuir_Data_untreated_nograph.png|200|thumb|Figure 2: Raw Data from the experiment]]
+
[[File:Langmuir_Data_untreated_nograph.png|200|thumb|Fig. 2 - Raw Data of the experiment]]
[[File:Langmuir_Data_untreated.png|200|thumb|Figure 3: Linear fit to the ion saturation region]]
+
[[File:Langmuir_Data_untreated.png|200|thumb|Fig. 3 - Linear fit to the ion saturation region]]
[[File:Langmuir_Data_treated.png|200|thumb|Figure 4: Fit to probe characteristic equation]]
+
[[File:Langmuir_Data_treated.png|200|thumb|Fig. 4 - Fit to probe characteristic equation]]
 +
 
 +
The Langmuir probe uses one or more electrodes placed inside the plasma to measure the current collected while applying a voltage. This will provide the I-V characteristic for that plasma. Normally a triangular or sawtooth wave is used to get equiseparated points.
 +
 
 +
When the probe is electrically isolated, a plasma sheath is formed in the interface between the plasma and the probe. To compensate for the higher mobility of the electrons, the probe will attain a floating potential,\(V_f\), negative in relation to the plasma potential itself, \(V_p\). The density at the sheath entrance is roughly half of the density far away from the probe.
  
The Langmuir probe uses one or more electrodes which are placed inside the plasma, in this case one of the filaments of the head. Measuring the current collected by the probe while applying a triangular wave, will provide the I-V characteristic that will be analysed later.
+
The probe voltage, \(V_s\) is sweeped in relation to the ground winding filament, using a variable voltage source. If the polarization of the probe compared to the plasma potential is negative enough, all the electrons will be repelled and the ion flux to the probe is independent of the potential applied. In a totally ionized plasma, this ion saturation current is described by the following expression:
 +
<math> i^+_{sat} = j^+_{sat} A_s = \frac{e \, n \, c_s \, A_s}{2} </math>
  
When the probe is electrically isolated, a plasma sheath is formed in the interface between the plasma and the probe. To compensate for the higher mobility of the electrons, the probe will attain a floating potential,\(V_f\), negative with respect to the plasma potential, \(V_p\). The density at the sheath entrance is roughly half of the density away from the probe.
+
Being \(j^+_{sat}\) the ion current density, \(A_s\) the contact surface of the probe, \(e\) the electron charge, \(n\) the ion density in the plasma, and \(c_s\) the ion sound speed.
  
The probe voltage, \(V_s\), has been changed with respect to the ground set by the winding filament using a variable voltage source. If the polarization of the probe, compared to the plasma, is negative enough, all the electrons will be repelled  nd the ion flux to the probe is independent of the tension applied. In a totally ionized plasma, this ion saturation current is described by the following expression:
+
When the polarization applied to the probe increases, the voltage drop in the sheath is reduced and the electrons will be able to reach the probe. Taking a Maxwell distribution of the speed of the electrons, the relation between current and tension will be:
\[ i^+_{sat} = j^+_{sat} A_s = e \, n \, c_s \, A_s \]
+
<math> i = -i^+_{sat} \left( 1- e^{\dfrac{e}{k T_e} (V_s - V_f)} \right) </math>
  
Where \(j^+_{sat}\) is the current density, \(A_s\) is the contact surface of the probe, \(e\) is the electron charge, \(n\) is the ion density in the plasma, \(c_s\) is the ion sound speed.
+
Being \(T_e\) the electron temperature. This expression assumes that there is only one probe and that it is non-perturbative.
  
If we polarise the probe positively, the voltage drop in the sheath is reduced and electrons will be able to reach the probe. Taking a Maxwell distribution for the speed of the electrons, the relation between current and tension will be:
+
From the last equation it is possible to extract an estimate of the floating potential, \(V_f\). This is done by taking the value at which the characteristic crosses zero.
\[ i = i^+_{sat} \left( 1- e^{\dfrac{e}{k T_e} (V_s - V_f)} \right) \]
 
  
Where \(T_e\) is the electron temperature. This expression assumes that there is only one probe and that it is non-perturbative.
+
It will also be easily seen that the data does not follow the regular characteristic on the ion saturation side, it should be constant instead of having a slope. This has to do with the fact that the sheath thickness expands with the applied voltage. To correct this, a linear function is fitted on this side and the slope is subtracted (as seen in Figure 3). Alternatively you can include this contribution in the equation:
  
From the obtainend it is possible to extract an estimate for the floating potential, \(V_f\). This is done by taking the value at which the characteristic crosses zero.
+
<math> i = -i^+_{sat} \left( 1- \alpha\ (V_s-V_f) - e^{\dfrac{e}{k T_e} (V_s - V_f)} \right) </math>
  
It will also be easily seen that the data does not follow a the regular characteristic on the ion saturation side, it should be constant instead of having a slope. This has to do with the fact that the probe is cylindrical. To correct for this, a linear function is fitted on this side and the slope is subtracted (as seen in Figure 3).
+
However, when the goal is to get the density, another correction must be done, which corresponds to adding the value of current at the point where the current is completely due to ions, namely to voltage values much lower than the floating potential (in this case, use two times the floating potential, hence the need of the initial estimate).
  
But, in doing this we impose a condition of cold ions, which means that there will be no ion saturation current, since the ions aren't moving. However, we want to get the density, so we have to make another correction, which corresponds to add the value of current in point where we know the current is completely due to ions, namely to values voltage much lower than the floating potential (in this case we use two times the floating potential, hence the need for our initial estimate).
+
After that, the experimental data has to be adjusted to the equation presented above (as seen in Figure 4). From the fit, the \(T_e\) can be extracted as well as the ion saturation current, \(i^+_{sat}\), and the floating potential, \(V_f\), which should have a value close to the one we determined empirically.
  
After that, the experimental data has to be adjusted to the equation presented in above (as seen in Figure 4). From the fit the \(T_e\) can be extracted as well as the ion saturation current, \(i^+_{sat}\) and the floating potential \(V_f\) which should have a value close to the one we determined empirically.
+
By knowing the area of the probe and the speed of sound in the ions of the plasma, you can determine the electron density in the plasma.
 +
Since \(c_s = \sqrt{\dfrac{k T_e}{M}}\) and given that the probe is \(10mm\) and has a diameter of \(0.2mm\), you can use the \(i^+_{sat}\) to determine the density.
  
By knowing the area of the probe and the speed of sound in the ions of the plasma, one can determine the electron density in the plasma.
+
=Bibliography=
Since \(c_s = \sqrt{\dfrac{k T_e}{M}}\), and the fact that the probe is \(1cm\) and has a diameter of \(0.2mm\) we can use the \(i^+_{sat}\) to determine the density.
+
*Principles of Plasma Diagnostics, Hutchinson, Cambridge.
  
 
=Links=
 
=Links=
<!--
+
*[[Sonda_de_Langmuir | Portuguese version (Versão em Português)]]
*[[Determinação da Constante Dieléctrica num Condensador Cilíndrico | Portuguese Version (Versão em Português)]]
 
*[http://en.wikipedia.org/wiki/Polystyrene Wikipedia page for Polystyrene]
 
-->
 

Revision as of 23:37, 17 May 2021

Description of the Experiment

Fig. 1 - Photo of the experimental setup

Plasmas have different characteristics from other states of matter and many diagnostic tools have been developed in order to measure their parameters. This experiment allows the measurement of some of these characteristics using one of the most simple methods, an electrical probe, the Langmuir probe. It consists of a thin filament made of a conductive material placed inside the plasma, which either attracts or repels the electrons in the plasma in accordance with its polarization. Measuring the probe I-V characteristic, that is the relationship between the polarization voltage and the respective collected current, you can determine the electron temperature and density of the plasma.

Links

  • Video: rtsp://elabmc.ist.utl.pt/langmuir.sdp
  • Laboratory: Advanced in e-lab.ist.utl.pt[1]
  • Control room: Langmuir Probe
  • Level: ****

Videos

Experimental Apparatus

The vacuum chamber consists of a Pyrex cylinder on which a glow discharge is produced by RF power. A tungsten filament with a diameter of 0.2mm is placed in the middle of the chamber partially isolated by an alumina cover and only a 10mm tip is exposed to the plasma, acting as a Langmuir probe. This central filament has a ground reference around the probe with approximately 200mm, in order to be much longer than the probe itself.

The RF generator (3.5W, 50kHz, 1kV) that ionizes the gas and generates the plasma is connected to two parallel molybdenum plates similar to a capacitor. NOTE: Gas pressure correction should be computed according Pfeifer's vacuum gauge PPT100 spreadsheet.

Protocol

Determination of the electron temperature and density

Fig. 2 - Raw Data of the experiment
Fig. 3 - Linear fit to the ion saturation region
Fig. 4 - Fit to probe characteristic equation

The Langmuir probe uses one or more electrodes placed inside the plasma to measure the current collected while applying a voltage. This will provide the I-V characteristic for that plasma. Normally a triangular or sawtooth wave is used to get equiseparated points.

When the probe is electrically isolated, a plasma sheath is formed in the interface between the plasma and the probe. To compensate for the higher mobility of the electrons, the probe will attain a floating potential,\(V_f\), negative in relation to the plasma potential itself, \(V_p\). The density at the sheath entrance is roughly half of the density far away from the probe.

The probe voltage, \(V_s\) is sweeped in relation to the ground winding filament, using a variable voltage source. If the polarization of the probe compared to the plasma potential is negative enough, all the electrons will be repelled and the ion flux to the probe is independent of the potential applied. In a totally ionized plasma, this ion saturation current is described by the following expression: [math] i^+_{sat} = j^+_{sat} A_s = \frac{e \, n \, c_s \, A_s}{2} [/math]

Being \(j^+_{sat}\) the ion current density, \(A_s\) the contact surface of the probe, \(e\) the electron charge, \(n\) the ion density in the plasma, and \(c_s\) the ion sound speed.

When the polarization applied to the probe increases, the voltage drop in the sheath is reduced and the electrons will be able to reach the probe. Taking a Maxwell distribution of the speed of the electrons, the relation between current and tension will be: [math] i = -i^+_{sat} \left( 1- e^{\dfrac{e}{k T_e} (V_s - V_f)} \right) [/math]

Being \(T_e\) the electron temperature. This expression assumes that there is only one probe and that it is non-perturbative.

From the last equation it is possible to extract an estimate of the floating potential, \(V_f\). This is done by taking the value at which the characteristic crosses zero.

It will also be easily seen that the data does not follow the regular characteristic on the ion saturation side, it should be constant instead of having a slope. This has to do with the fact that the sheath thickness expands with the applied voltage. To correct this, a linear function is fitted on this side and the slope is subtracted (as seen in Figure 3). Alternatively you can include this contribution in the equation:

[math] i = -i^+_{sat} \left( 1- \alpha\ (V_s-V_f) - e^{\dfrac{e}{k T_e} (V_s - V_f)} \right) [/math]

However, when the goal is to get the density, another correction must be done, which corresponds to adding the value of current at the point where the current is completely due to ions, namely to voltage values much lower than the floating potential (in this case, use two times the floating potential, hence the need of the initial estimate).

After that, the experimental data has to be adjusted to the equation presented above (as seen in Figure 4). From the fit, the \(T_e\) can be extracted as well as the ion saturation current, \(i^+_{sat}\), and the floating potential, \(V_f\), which should have a value close to the one we determined empirically.

By knowing the area of the probe and the speed of sound in the ions of the plasma, you can determine the electron density in the plasma. Since \(c_s = \sqrt{\dfrac{k T_e}{M}}\) and given that the probe is \(10mm\) and has a diameter of \(0.2mm\), you can use the \(i^+_{sat}\) to determine the density.

Bibliography

  • Principles of Plasma Diagnostics, Hutchinson, Cambridge.